63
NotesNotes
NotesNotes
Notes If no turbidity calibration has been performed or if calibration was
cleared using the “Restore factory calib.” option the offset and slope
values are set to default, and the message ”Factory calibration” is dis-
played. Press <ESC> to return to the previous screen.
Temperature
From the “GLP” menu select the “Temperature”
option.
Data regarding the last temperature calibration
will be displayed: calibrated point, time and date.
Use the arrow keys to scroll through the stored
data for the last 5 calibrations.
NotesNotes
NotesNotes
Notes If no user temperature calibration has been performed or if calibration
was cleared using the “Restore factory calib.” option the offset value is
set to default, and the message ”Factory calibration” is displayed. Press
<ESC> to return to the previous screen.
Atmospheric Pressure
From the “GLP” menu select “Atm. pressure”.
Data regarding the last atmospheric pressure
calibration will be displayed: custom calibration
point, time and date.
Use the arrow keys to scroll through the stored
data for the last 5 calibrations.
NotesNotes
NotesNotes
Notes If no atmospheric pressure calibration has been performed or if cali-
bration was cleared using the “Restore factory calib.” option the offset
value is set to default, and the message ”Factory calibration” is dis-
played. Press <ESC> to return to the previous screen.