Hanna-instruments HI 9812-5N Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Instruments de mesure Hanna-instruments HI 9812-5N. Hanna Instruments HI 9812-5N User Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 16
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 0
1
Instruction Manual
HI 9812-5
Portable
pH/EC/TDS/°C Meter
www.hannainst.com
Vue de la page 0
1 2 3 4 5 6 ... 15 16

Résumé du contenu

Page 1 - Portable

1Instruction ManualHI 9812-5PortablepH/EC/TDS/°C Meterwww.hannainst.com

Page 2

10Accessories needed:• Use HI 70031 (1413 µS/cm) EC calibration solution orHI 70032 (1382 ppm) TDS calibration solution.Note: The conversion between E

Page 3

11The TDS value in aqueous solutions is directly proportional to theconductivity. The ratio between the two parameters depends on thesolution.The inst

Page 4 - FUNCTIONAL DESCRIPTION

12These meters are powered by a 9V battery that is located on the rearof the instrument.When battery symbol is empty a low battery condition is indica

Page 5 - SPECIFICATIONS

13PERIODIC MAINTENANCEInspect the probe and the cable. The cable used for the connection tothe meter must be intact and there must be no points of bro

Page 6 - OPERATIONAL GUIDE

14PROBESHI 1285-5 Combination, amplified pH/EC/TDS/temperature probewith built-in temperature sensor, 8-pin DIN connectorand 1 m (3.3') cablepH B

Page 7

15RECOMMENDATIONS FOR USERSBefore using these products, make sure they are entirely suitable forthe environment in which they are used.Operation of th

Page 8

16Printed in EUROPE(ROMANIA)MAN9812 09/10Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USATechnical Support f

Page 9

2WARRANTY ... 2PRELIMINARY EXAMINATION ...

Page 10 - Accessories needed:

3All rights are reserved. Reproduction in whole or in part is prohibited without the writtenconsent of the copyright owner.PRELIMINARY EXAMINATIONPREL

Page 11 - EC/TDS CONVERSION FACTOR

41) 8-pin DIN connector for probe2) Liquid Crystal Display3) pH range selection key4) µS/cm (EC) range selection key5) ppm (TDS) selection key6) pH of

Page 12 - BATTERY REPLACEMENT

5egnaRHp0.41ot0.00991ot0µ mc/Smpp0991ot0C°06ot0noituloseRHp1.001µ mc/Smpp01C°1ycaruccA)F°86/C°02@(Hp1.0±.s.f%2±µ mc/Smpp.s.f%2±C°1±noitaiveDCMElacipyT

Page 13 - PROBE MAINTENANCE

6INITIAL PREPARATIONEach meter is supplied complete with a 9V battery. Remove thebattery compartment cover on the back of the meter and install thebat

Page 14 - ACCESSORIES

7• Select the appropriate measurement range (EC or TDS).• Wait a couple of minutes for the temperature sensor to reachthermal equilibrium. The display

Page 15

8For greatest accuracy, frequent calibration of the instrument isrecommended. The instrument should be recalibrated for pH:a) Whenever the electrode i

Page 16 - Printed in EUROPE

9For temperature compensation during calibration, please refer to thefollowing chart.For instance, if the buffer temperature is 25°C, the display shou

Commentaires sur ces manuels

Pas de commentaire