7.9 ATMOSPHERIC PRESSURE CALIBRATION
Place HI 9829
in a wind-free area and choose “Custom pressure” to perform a
user calibration or “Restore factory calib”.
NoteNote
NoteNote
Note “Custom pressure” procedure requires a reference barometer.
Select the “Atm. pressure” from the “Calibration” menu.
Select the “Custom pressure” option.
Using the keypad, insert the numeric value that agrees with the reference meter
and then press <Accept> to confirm.
The stability counter will count down and the message “Ready” and “Confirm
will be displayed. Press <Confirm> to store the calibration point.
After confirmation, the following messages are displayed: “Storing” and “Cali-
bration completed”.
Press <Measure> to return to the measurement screen.
Press <OK> to return to the “Calibration” menu.
To restore the factory calibration, select “Restore factory calib.” in the “Pressure
calibration” menu and press <Select>.
After confirmation, the following messages are displayed “Storing” and “Cali-
bration completed”.
Press <OK> to return to the “Calibration” menu.
Press <Measure> to return to the measurement screen.
To restore the factory calibration, select the corresponding option in the
“Temperature calib.” menu and then press <Select>.