SALINITY CALIBRATIONSALINITY CALIBRATION
SALINITY CALIBRATIONSALINITY CALIBRATION
SALINITY CALIBRATION
Salinity calibration is a one-point procedure at 100.0% NaCl. Use the HI 7037L calibration solution
(sea water solution) as a 100% NaCl standard solution.
To enter salinity calibration:
Set the meter for salinity range;
Select the
Percent Scale
(see
Salinity Setup
section);
Rinse the probe with some of the calibration solution or deionized water;
Immerse the probe into HI 7037L solution. The sleeve holes must be completely submerged. Tap
the probe repeatedly to remove any air bubbles that may be trapped inside the sleeve.
Enter in calibration mode by pressing
;
Wait to stabilize;
Press
to finish salinity calibration or press to cancel calibration.