Temperature Compensation
The user can choose from the following options:
Linear -
the meter will compensate automatically the conductivity using the following formula:
where:
C
ref
- conductivity at reference temperature
C
i
- measured conductivity (uncompensated)
- compensation coefficient
T - temperature
ref - reference temperature
Non-Linear -
recommended for measuring the conductivity of the natural water in accordance with the
compensation table on the next page.
Disabled -
the meter will display the conductivity with no temperature compensation.
To set the
temperature compensation
mode:
Use
or to highlight the
Temperature
Compensation
option.
Press
and then use or to select
Linear, Non-Linear
or
Disabled
option.
Press
to confirm your selection or press to cancel
operation.