10
For each measurement mode (Conductivity, Resistivity, TDS or Salinity) the following display configurations
are available: Basic, Good Laboratory Practice (GLP), Graph and Log History.
Basic
Accessing this option, the measured value and its units are dis-
played on the LCD, along with the temperature value, temperature
compensation mode, and minimal GLP data.
To choose the Basic displaying mode:
Press
while in
Measure
mode. The “Choose Display
Configuration” message will be displayed in the Reminder
messages area.
Press
. The instrument will display the basic informa-
tion for the selected measurement mode.
GLP
Accessing this option, a detailed GLP data will be displayed on the
LCD for conductivity and salinity (Percent Scale only)
measure
modes: Last Calibration Date and Time, Cell Constant, Probe
Offset, Reference Temperature, Compensation Coefficient, Tem-
perature Compensation.
To access the GLP displaying mode:
Press
while in
Measure
mode. The “Choose Display
Configuration” message will be displayed in the Reminder
messages area.
Press
. The instrument will display the detailed GLP
data.