METHODS
5.5.12 Result Unit
The titrator provides the results based on the selected units.
5.6 Printing
To print method parameters, press from the main screen, then .
If no printer is connected to the dedicated socket, or if the printer is offline, an error message
will appear on the display (see Section
8.5.3, Connecting a Printer
for information about
connecting a printer to the titrator).
5.5.11.4 End Point Stability Time
This screen allows users to set the time period in which the electrode potential must remain
stable.
This setting is in accordance with the
mV End Point
termination criterion.