2-9
SETUP
Figure 3
2.3.4.3 Titration Vessel
The titration takes place in a sealed titration vessel. The titration vessel can also be referred
to as a reaction vessel, titration cell or reaction cell.
The primary design features of the HI 904 titration vessel include the following:
Durability, easy to use, clean and maintain.
All glass cells are manufactured with ground-glass joints for ultra-low water vapor
permeability and high chemical resistivity to Karl Fischer reagents.
A sample port with open-top screw cap for easy septum replacement.
A desiccant cartridge containing molecular sieves to dry the ambient air which
enters the cell as reagent and sample are added/removed from the titration vessel.
Electrode port for generators with and without diaphragm.
To attach the titration vessel, see Figure 3 and follow the steps below:
Align the vessel support (D) with the base plate and attach by rotating clockwise.
Insert the support adapter (C) onto the vessel support (D) while aligning the notches.
If the o-ring (B) is not installed, insert it into the inner groove of the support adapter (C).
Lower titration vessel into the vessel support (D) by gently pushing it through the
support adapter (C).