Hanna-instruments HI 83207-01 Manuel d'utilisateur Page 18

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 68
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 17
18
Aluminum
Add the content of one packet of HI 93712C-0 Bleaching
powder to one of the two cuvettes. Replace the cap and shake
vigorously until completely dissolved. This is the blank.
Place the blank into the holder and close the lid.
Press Timer and the display will show the countdown prior to
zeroing the blank. Alternatively wait for 15 minutes and then press
Zero. The display will show “-0.0-” when the meter is zeroed and
ready for measurement.
Remove the blank and insert the other cuvette into the instrument.
Press the Read key and the meter will perform the reading. The instrument displays the results in
mg/L of aluminum.
Press or to access the second level functions.
Press the Chem Frm key to convert the result in mg/L of Al
2
O
3
.
Press or to return to the measurement screen.
INTERFERENCES
Interference may be caused by:
Iron above 20 mg/L, Alkalinity above 1000 mg/L, Phosphate above 50 mg/L, Fluoride must be absent.