Hanna-instruments HI 83203 Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Instruments de mesure Hanna-instruments HI 83203. Hanna Instruments HI 83203 User Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer

Résumé du contenu

Page 1 - HI 83203

1Instruction ManualHI 83203Multiparameter BenchPhotometerfor Aquaculturewww.hannainst.com

Page 2 - TABLE OF CONTENTS

10HEALTH & SAFETY• The chemicals contained in the reagent kits may be hazardous if improperly handled.• Read the Material Safety Data Sheet (MSDS)

Page 3 - GENERAL DESCRIPTION

11POWER CONNECTION AND BATTERY MANAGEMENTThe meter can be powered from an AC/DC adapter (included) or from the built-in rechargeable battery.Note: Alw

Page 4 - PRECISION AND ACCURACY

12DATA MANAGEMENTThe instrument features a data log function to help you keep track of all your analysis. The data log canhold 200 individual measurem

Page 5

13In the Setup mode the instrument’s parameters can be changed. Some parameters affect the measuringsequence and others are general parameters that ch

Page 6 - FUNCTIONAL DESCRIPTION

14Date / TimeThis option is used to set the instrument’s date and time.Press the Modify key to change the date/time.Press the W X functional keys to h

Page 7 - KEYPAD DESCRIPTION

15HI 83203 offers an interactive contextual help mode that assists the user at any time.To access the help screens press HELP.The instrument will disp

Page 8

16Ammonia MRSPECIFICATIONSRange 0.00 to 10.00 mg/LResolution 0.01 mg/LAccuracy ±0.05 mg/L ±5% of reading at 25 °CTypical EMC ±0.01 mg/LDeviationLight

Page 9 - USING CUVETTES

17• Remove the cuvette.• Add 4 drops of HI 93715A-0 First reagent (6 drops forseawater analysis). Replace the cap and mix the solution.• Add 4 drops o

Page 10 - HEALTH & SAFETY

18Ammonia LRSPECIFICATIONSRange 0.00 to 3.00 mg/LResolution 0.01 mg/LAccuracy ±0.04 mg/L ±4% of reading at 25 °CTypical EMC ±0.01 mg/LDeviationLight S

Page 11 - METHOD REFERENCE TABLE

19Ammonia LR• Remove the cuvette.• Add 4 drops of HI 93700A-0 First reagent (6 drops forseawater analysis). Replace the cap and mix the solution.• Add

Page 12 - METHOD SELECTION

2Dear Customer,Thank you for choosing a Hanna product. Please read this instruction manual carefully before using theinstrument. This manual will prov

Page 13 - SETUPSETUP

20FREE CHLORINEFREE CHLORINEFREE CHLORINEFREE CHLORINEFREE CHLORINESPECIFICATIONSRange 0.00 to 2.50 mg/LResolution 0.01 mg/LAccuracy ±0.03 mg/L ±3% of

Page 14

21Free Chlorine• Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.• Remove the cuvette.Powder reagen

Page 15 - HELP MODE

22• Reinsert the cuvette into the instrument.• Press Read to start the reading. The instrument displays the results in mg/L of free chlorine.INTERFER

Page 16 - AMMONIA MEDIUM RANGE

23Total ChlorineTOTAL CHLORINESPECIFICATIONSRange 0.00 to 3.50 mg/LResolution 0.01 mg/LAccuracy ±0.03 mg/L ±3% of reading at 25 °CTypical EMC ±

Page 17 - INTERFERENCES

24Total Chlorine• Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.• Remove the cuvette.Powder reage

Page 18 - AMMONIA LOW RANGE

25• Press Timer and the display will show the countdown prior to the measurement or, alternatively, waitfor 2 minutes and 30 seconds and press Read. W

Page 19

26Copper HRCOPPER HIGH RANGESPECIFICATIONSRange 0.00 to 5.00 mg/LResolution 0.01 mg/LAccuracy ±0.02 mg/L ±4% of reading at 25 °CTypical EMC ±0.01 mg/L

Page 20 - FREE CHLORINE

27Copper HR• Remove the cuvette.• Add the content of one packet of HI 93702-0 Bicinchoninate.Replace the cap and shake gently for about 15 seconds.• R

Page 21 - Free Chlorine

28Copper LRCOPPER LOW RANGESPECIFICATIONSRange 0 to 1000 μg/LResolution 1 μg/LAccuracy ±10 μg/L ±5% of reading at 25 °CTypical EMC ±1 μg/LDeviationLig

Page 22

29Copper LR• Remove the cuvette.• Add the content of one packet of HI 95747-0 Bicinchoninate.Replace the cap and shake gently for about 15 seconds.• R

Page 23 - TOTAL CHLORINE

3PRELIMINARY EXAMINATIONPlease examine this product carefully. Make sure that the instrument is not damaged. If any damageoccurred during shipment, pl

Page 24 - Total Chlorine

30NitrateSPECIFICATIONSRange 0.0 to 30.0 mg/LResolution 0.1 mg/LAccuracy ±0.5 mg/L ±10% of reading at 25 °CTypical EMC ±0.1 mg/LDeviationLight Source

Page 25

31Nitrate• Replace the cap and immediately shake vigorously upand down for exactly 10 seconds. Continue to mix byinverting the cuvette gently for 50 s

Page 26 - COPPER HIGH RANGE

32Nitrite HRNITRITE HIGH RANGESPECIFICATIONSRange 0 to 150 mg/LResolution 1 mg/LAccuracy ±4 mg/L ±4% of reading at 25 °CTypical EMC ±1 mg/LDeviationLi

Page 27

33Nitrite HR• Add the content of one packet of HI 93708-0 reagent.Replace the cap and shake gently until completely dissolved.• Reinsert the cuvette i

Page 28 - COPPER LOW RANGE

34SPECIFICATIONSRange 0.00 to 1.15 mg/LResolution 0.01 mg/LAccuracy ±0.06 mg/L ±4% of reading at 25 °CTypical EMC ±0.01 mg/LDeviationLight Source Tung

Page 29

35• Reinsert the cuvette into the instrument.• Press Timer and the display will show the countdown prior to the measurement or, alternatively, waitfor

Page 30 - MEASUREMENT PROCEDURE

36Dissolved OxygenSPECIFICATIONSRange 0.0 to 10.0 mg/LResolution 0.1 mg/LAccuracy ±0.4 mg/L ±3% of reading at 25 °CTypical EMC ± 0.1 mg/LDeviationLigh

Page 31

37Dissolved Oxygen• Let the sample stand and the flocculent agent will start to settle.• After approximately 2 minutes, when the upper half of the bot

Page 32 - NITRITE HIGH RANGE

38pHpHSPECIFICATIONSRange 6.5 to 8.5 pHResolution 0.1 pHAccuracy ±0.1 pH at 25 °CTypical EMC ±0.1 pHDeviationLight Source Tungsten lamp with narrow ba

Page 33 - Nitrite HR

39pH• Remove the cuvette and add 5 drops of HI 93710-0 PhenolRed Indicator. Replace the cap and mix the solution.• Reinsert the cuvette into the instr

Page 34 - NITRITE LOW RANGE

4PRINCIPLE OF OPERATIONAbsorption of light is a typical phenomenon of interaction between electromagnetic radiation and matter.When a light beam cross

Page 35

40Phosphate HRPHOSPHATE HIGH RANGESPECIFICATIONSRange 0.0 to 30.0 mg/LResolution 0.1 mg/LAccuracy ±1 mg/L ±4% of reading at 25 °CTypical EMC Dev. ±0.1

Page 36 - DISSOLVED OXYGEN

41Phosphate HR• Add the content of one packet of HI 93717B-0 PhosphateHR Reagent B to the cuvette. Replace the cap and shakegently until completely di

Page 37

42Phosphate LRPHOSPHATE LOW RANGESPECIFICATIONSRange 0.00 to 2.50 mg/LResolution 0.01 mg/LAccuracy ±0.04 mg/L ±4% of reading at 25 °CTypical EMC Dev.

Page 38

43Phosphate LR• Reinsert the cuvette into the instrument.• Press Timer and the display will show the countdown prior to the measurement or, alternativ

Page 39

44The instrument shows clear warning messages when erroneous conditions appear and when measured values areoutside the expected range. These messages

Page 40 - PHOSPHATE HIGH RANGE

45Description Range MethodAmmonia MR 0.00 to 10.00 mg/ L NesslerAmmonia LR 0.00 to 3.00 mg/ L NesslerChlorine, Free 0.00 to 2.50 mg/ L DPDChlorine, To

Page 41

46OTHER ACCESSORIESHI 731318 Cloth for wiping cuvettes (4 pcs)HI 731321 glass cuvettes (4 pcs)HI 731325W new cap for cuvette (4 pcs)HI 740034 cap for

Page 42 - PHOSPHATE LOW RANGE

47WARRANTYAll Hanna Instruments meters are warranted for two years against defects in workmanship and materialswhen used for its intended purpose and

Page 43

48Printed in EUROPE(ROMANIA)Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USATechnical Support for CustomersTe

Page 44 - ERRORS AND WARNINGS

5Where:-log I/Io= Absorbance (A)Io= intensity of incident light beamI = intensity of light beam after absorptionελ= molar extinction coefficient at wa

Page 45 - STANDARD METHODS

6INSTRUMENT DESCRIPTION1) Open Cuvette Lid2) Indexing mark3) Cuvette point4) Liquid Crystal Display (LCD)5) Splash proof keypad6) ON/OFF power switch7

Page 46 - ACCESSORIES

7KEYPAD DESCRIPTIONThe keypad contains 8 direct keys and 3 functional keys with the following functions:Press to perform the function displayed above

Page 47 - HANNA LITERATURE

8(c) take out the syringe and clean the outside of the syringe tip. Be sure that no drops are hangingon the tip of the syringe, if so eliminate them.

Page 48 - MAN83203 04/10

9shake gently shake vigorouslyUSING CUVETTES• Proper mixing is very important for reproducibility of the measurements. The right way of mixing acuvett

Commentaires sur ces manuels

Pas de commentaire