The 3 options are correlated, so that each of the procedures will calibrate all the
3 ranges.
NoteNote
NoteNote
Note For correct EC readings, calibration should be performed using a stan-
dard solution with a conductivity value close to the sample to be mea-
sured.
After choosing the “Conductivity calibration” mode from the “Calibration” menu,
select the calibration type with the arrow keys and then press <OK>.
Conductivity
Select the “Conductivity” option and press <OK> to confirm.
Fill a beaker with a standard conductivity solution (see “Accessories” section for
choosing the proper HANNA solution).
Immerse the probe in the solution and wait for stable reading. The probe sleeve
must be inserted.
The main display shows the actual reading, while the secondary level displays
the standard value.
To change the standard value, press <Cal. point> and the list of available
standard values is displayed: 0 µS/cm, 84 µS/cm, 1413 µS/cm, 5.00 mS/cm,
12.88 mS/cm, 80.0 mS/cm and 111.8 mS/cm.
Press <Custom> to insert a custom value (temperature compensated value).
Choose <Resolution> to select the desired resolution.
When the reading is stable, press <Confirm> to store the calibration data.
After confirmation, the following messages are displayed: “Storing data on
probe, please wait...”, “Updating GLP data, please wait...” and “Calibration
completed”.
Press <OK> to return to the “Calibration” menu.
To return to the main menu, press ESC repeatedly.