6.6 ORP
From GLP data main menu select the “ORP” option and press OK.
All the information about the last ORP calibration appears: calibrated point,
time and date
In case of restoring factory calibration data, the
display will show the warning message “Factory
calibr. restored”.
Use the arrow keys to scroll the last 5 memo-
rized calibrations.
NONO
NONO
NO
TETE
TETE
TE If no ORP calibration has been performed, a warning message advise
the user. Press OK to return to the previous screen.
If the selected calibration is a factory calibration, the meter shows the
message “Factory calibration”.
6.7 TEMPERATURE
From GLP data main menu select the “Temperature” option and press OK.
All the information about the last temperature
calibration appears: calibrated point, time and
date.
Use the arrow keys to scroll the last 5 memo-
rized calibrations.
NONO
NONO
NO
TETE
TETE
TE If no temperature calibration has been performed, a warning message
advise the user. Press OK to return to the previous screen.
If the selected calibration is a factory calibration, the meter shows the
message “Factory calibration”.