Add the content of one packet of HI 93722-0 Cyanuric Acid
Reagent. Replace the cap and shake gently for about 10
seconds (dissolution is not complete).
Reinsert the cuvette into the instrument.
Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 45 seconds and press Read. When the timer ends the meter will perform the reading.
The instrument displays concentration in mg/L of cyanuric acid.
INTERFERENCES
Turbidity preexisting in the sample causes interference during measurement.
Cyanuric Acid