Hanna-instruments HI 3220 Manuel d'utilisateur Page 57

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 68
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 56
57
PREPARATION PROCEDURE
Remove the electrode protective cap.
DO NOT BE ALARMED IF ANY SALT DEPOSITS ARE PRESENT. This is normal
with electrodes and they will disappear when rinsed with water.
During transport tiny bubbles of air may have formed inside the glass bulb.
The electrode cannot function properly under these conditions. These bubbles
can be removed by "shaking down" the electrode as you would do with a glass
thermometer.
If the bulb and/or junction are dry, soak the electrode in HI 70300 Storage
Solution for at least one hour. Open the fill hole if refillable.
Not present in gel electrodes.
ELECTRODE CONDITIONINGELECTRODE CONDITIONING
ELECTRODE CONDITIONINGELECTRODE CONDITIONING
ELECTRODE CONDITIONING
& MAINTENANCE& MAINTENANCE
& MAINTENANCE& MAINTENANCE
& MAINTENANCE
Vue de la page 56
1 2 ... 52 53 54 55 56 57 58 59 60 61 62 ... 67 68

Commentaires sur ces manuels

Pas de commentaire