The mV reading, with sign and decimal point (7 chars)
The logged time, yymmddhhmmss (12 chars)
The calibration slope, with sign and decimal point
(7 chars) - not available for Rel mV range
The calibration offset, with sign and decimal point
(7 chars) - not available for ISE
Temperature probe presence (1 char)
Notes: “Err8” is sent if the instrument is not in measurement mode.
“Err6” is sent if the requested range is not available.
“Err4” is sent if the requested parameter is not available.
“Err3” is sent if the Log on demand is empty.
“Err9” is sent if the battery power is less than 30%.
Invalid commands will be ignored.