Default slope value (mV/decade):
–59.16 (monovalent cation) - ion charge is 1
59.16 (monovalent anion) - ion charge is –1
–29.58 (divalent cation) - ion charge is 2
29.58 (divalent anion) - ion charge is –2
100 - ion charge is “undF
Press CLR during calibration if you want to clear calibration
values. The instrument will display “CLR” message and will
return to measurement mode.
The calibration needs to be cleared when a new electrode is
attached to the meter.
The instrument will display “----” on the primary LCD if it is
not calibrated or after all calibrations are cleared.
Press RANGE to display the temperature reading on
the LCD during calibration.
ONE-POINT CALIBRATION
Press CAL after first calibration point was confirmed. The instrument
memorizes the one-point calibration information and returns to
measurement mode.
Press RANGE to display the temperature reading on
the LCD during calibration.