1Instruction ManualHI 96702Copper High Range ISMwww.hannainst.com
10Under range: A blinking “0.00” indicates that the sampleabsorbs less light than the zero reference. Check the procedureand make sure you use the sam
11The instructions listed below should be carefully followed during testing to ensure best accuracy.• Color or suspended matter in large amounts may c
12• Whenever the cuvette is placed into the measurement cell, itmust be dry outside, and completely free of fingerprints, oilor dirt. Wipe it thorough
13• When the beeper sounds briefly and theLCD displays dashes, the meter is ready.The blinking “ZERO” indicates that theinstrument needs to be zeroed
14• Remove the cuvette.• Add the content of one packet ofHI 93702-0 reagent. Replace thecap and shake gently.• Replace the cuvette into the cuvette ho
15VALIDATION PROCEDUREUse the validation procedure to ensure that the instrument is properly calibrated.Warning: Do not validate the instrument with a
16• Remove the cuvette.• Place the CAL CHECK™ StandardHI 96702-11 Cuvette B into the cuvetteholder and ensure that the notch on thecap is positioned s
17CALIBRATION PROCEDURENote: It is possible to interrupt calibration procedure at any time by pressing CAL CHECK orON/OFF keys.Warning: Do not calibra
18• After a few seconds the display willshow “-0.0-”. The meter is now zeroedand ready for calibration. The blinking“READ” asks for reading calibratio
19• Press GLP/ to edit the desired year(2009-2099). If the key is kept pressed,the year number is automaticallyincreased.• When the correct year has
2TABLE OF CONTENTSPRELIMINARY EXAMINATION ... 3GENERAL DE
20To display the calibration date:• Press GLP/ to enter GLP mode. Thecalibration month and day will appear onthe main display and the year on theseco
21• Press ZERO/CFM to restore the factorycalibration or press GLP/ again toabort factory calibration restore.• The instrument briefly indicates “donE
22• Turn the instrument off by pressing ON/OFF.• Turn the instrument upside down and remove the battery cover by turning it counterclockwise.• Extract
23Recommendations for UsersBefore using these products, make sure that they are entirely suitable for your specific application and for the environmen
24MAN96702 12/10Printed in EUROPE(ROMANIA)Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USATechnical Support
3PRELIMINARY EXAMINATIONPlease examine this product carefully. Make sure that the instrument is not damaged. If anydamage occurred during shipment, pl
4GENERAL DESCRIPTIONABBREVIATIONSThe HI 96702 is an auto diagnostic portable microprocessor meter that benefits from Hanna’syears of experience as a m
5SPECIFICATIONSPRECISION AND ACCURACYRange 0.00 to 5.00 mg/LResolution 0.01 mg/LAccuracy ±0.02 mg/L ±4% of reading @ 25°CTypical EMC Deviation ±0.01 m
6PRINCIPLE OF OPERATIONAbsorption of Light is a typical phenomenon of interaction between electromagnetic radiation andmatter. When a light beam cross
71) GLP/ key2) CAL CHECK key3) ZERO/CFM key4) READ/ /TIMER key5) ON/OFF key6) Liquid Crystal Display (LCD)7) Cuvette alignment indicator8) Cuvette ho
810852973461KEYPAD DESCRIPTION• ON/OFF: to turn the meter on and off.• ZERO/CFM: this is a bi-functional key. Just press to zero the meter prior to me
9ERRORS AND WARNINGSThe instrument shows clear messages when erroneous condition appears. Messages are also displayedwhen the obtained values are outs