Instruction ManualHI 9835WaterproofEC/TDS/NaCl/°C Meterfor Lab and Fieldwww.hannainst.com
1918CE DECLARATION OF CONFORMITYWARRANTYAll Hanna Instruments meters are warranted for two yearsagainst defects in workmanship and materials when used
MAN9835R2 09/05SALES AND TECHNICAL SERVICE CONTACTSAustralia:Tel. (03) 9769.0666 • Fax (03) 9769.0699China:Tel. (10) 88570068 • Fax (10) 88570060
32Remove the instrument from the packing material and examine itcarefully to make sure that no damage has occurred during shipping.If there is any dam
54SPECIFICATIONSRange EC 0.00 to 29.99 μS/cm(Autoranging) 30.0 to 299.9 μS/cm300 to 2999 μS/cm3.00 to 29.99 mS/cm30.0 to 200.0 mS/cmup to 500.0 mS/cm
76EC Calibration 1 point, with 6 memorized buffers84,1413,5000,12880,80000,111800 μS/cmNaCl Calibration 1 point, with HI 7037 buffer (optional)Tempera
98EC/TDS CALIBRATIONEC/TDS calibration is a single point procedure. Selectable calibrationpoints are: 0.0, 84.0 μS/cm, 1413 μS/cm, 5.00 mS/cm, 12.88mS
1110TEMPERATURE CALIBRATION (for technical personnel only)The temperature calibration is a 2 point procedure, at 0 and 50°C.• Immerse the probe in a 0
1312The setup mode is used to view or change the instrument parameters.To enter setup press ALT+FNC when themeter is in measurement mode."Set&quo
1514PROBE MAINTENANCEAfter measurements, rinse the probe with clean water. If a morethorough cleaning is required, remove the probe sleeve and clean t
1716OTHER ACCESSORIESHI 76309 Stainless steel 4-ring conductivity probe with built-in temperature sensor and 1 m (3.3') cable.HI 76310 Platinum