14
Remove the cap.
Add the content of one packet of
HI 96770C-0 Silica HR reagent C and shake
vigorously until complete dissolution.
Replace the cuvet into the cuvet holder and
ensure that the notch on the cap is
positioned securely into the groove.
Wait for 2 minutes and then press
READ READ
READ READ
READ
/TIMER/TIMER
/TIMER/TIMER
/TIMER.
The instrument status indicator will appear
depending on the measurement phase.
The instrument directly displays
concentration in mg/L of silica (SiO
2
) on the
LCD.
To convert the reading to mg/L of silicon
(Si) multiply the reading by 0.467.