When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
The blinking “ZERO” indicates that the
instrument needs to be zeroed first.
Fill a graduated beaker with 50 mL of
sample.
Add 0.5 mL of HI 93719A-0 Magnesium
indicator solution, then swirl to mix.
In the same beaker add 0.5 mL of HI
93719B-0 Alkali solution for
Magnesium, then swirl to mix.
Fill two cuvettes with 10 mL of sample
each.
Add 1 drop of HI 93719C-0 EDTA solution
to one cuvette, replace the cap and swirl
the solution. This is the blank.
Add 1 drop of HI 93719D-0 EGTA solution
to the second cuvette, replace the cap
and swirl the solution. This is the sample.
Place the blank into the cuvette holder
and ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.
After a few seconds, the display will
show “-0.0-”. The meter is now zeroed
and ready for measurement.