Hanna-instruments HI 96713 Manuel d'utilisateur Page 13

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 24
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 12
13
Place the cuvette into the cuvette holder
and ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.
After a few seconds, the display will show
“-0.0-”. The meter is now zeroed and
ready for measurement. Remove the cuvette.
Add one packet of the HI 93713 reagent.
Replace the cap and shake gently (for
about 2 minutes) until the powder is
completely dissolved.
Replace the cuvette into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press and hold READ/ /TIMER for three
seconds. The display will show the
countdown prior to measurement. An
audible “beep” indicates the end of the
countdown period.
Vue de la page 12
1 2 ... 8 9 10 11 12 13 14 15 16 17 18 ... 23 24

Commentaires sur ces manuels

Pas de commentaire