1Instruction ManualHI 96709CManganese High Range ISMwww.hannainst.com
10Under range: A blinking “0” indicates that the sample absorbsless light than the zero reference. Check the procedure and makesure you use the same c
11The instructions listed below should be carefully followed during testing to ensure best accuracy.• Color or suspended matter in large amounts may c
12• Whenever the cuvette is placed into the measurement cell, itmust be dry outside, and completely free of fingerprints, oilor dirt. Wipe it thorough
13• When the beeper sounds briefly and theLCD displays dashes, the meter is ready.The blinking “ZERO” indicates that theinstrument needs to be zeroed
14• Remove the cuvette.• Add the content of one packet ofHI 93709A-0 Manganese HR ReagentA, replace the cap and shake gently thecuvette for 2 minutes.
15VALIDATION PROCEDURECalcium above 700 mg/LChloride above 70000 mg/LIron above 5 mg/LMagnesium above 100000 mg/LUse the validation procedure to ensur
16• After a few seconds, the display will show“-0.0-”. The meter is now zeroed and readyfor validation.• Remove the cuvette.• Place the CAL CHECKTM St
17CALIBRATION PROCEDURENote: It is possible to interrupt calibration procedure at any time by pressing CAL CHECK orON/OFF keys.Warning: Do not calibra
18• After a few seconds the display will show“-0.0-”. The meter is now zeroed and readyfor calibration. The blinking “READ” asks forreading calibratio
19• Press GLP/ to edit the desired year(2000-2099). If the key is kept pressed,the year number is automaticallyincreased.• Press GLP/ to edit the desi
2TABLE OF CONTENTSPRELIMINARY EXAMINATION ... 3GENERAL DESC
20To display the calibration date:• Press GLP/ to enter GLP mode. Thecalibration month and day will appear onthe main display and the year on theseco
21• Press ZERO/CFM to restore the factorycalibration or press GLP/ again toabort factory calibration restore.• Press READ/ /TIMER to enter in thefact
22• Turn the instrument off by pressing ON/OFF.• Turn the instrument upside down and remove the battery cover by turning it counterclockwise.• Extract
23Recommendations for UsersBefore using these products, make sure that they are entirely suitable for your specific application and for the environmen
24MAN96709 12/10Printed in EUROPE(ROMANIA)Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USATechnical Support
3PRELIMINARY EXAMINATIONPlease examine this product carefully. Make sure that the instrument is not damaged. If anydamage occurred during shipment, pl
4GENERAL DESCRIPTIONABBREVIATIONSThe HI 96709 is an auto diagnostic portable microprocessor meter that benefits from Hanna’syears of experience as a m
5SPECIFICATIONSPRECISION AND ACCURACYRange 0.0 to 20.0 mg/LResolution 0.1 mg/LAccuracy ±0.2 mg/L ±3% of reading @ 25°CTypical EMC Deviation ±0.1mg/LLi
6PRINCIPLE OF OPERATIONAbsorption of Light is a typical phenomenon of interaction between electromagnetic radiation andmatter. When a light beam cross
71) GLP/ key2) CAL CHECK key3) ZERO/CFM key4) READ/ /TIMER key5) ON/OFF key6) Liquid Crystal Display (LCD)7) Cuvette alignment indicator8) Cuvette h
810852973461KEYPAD DESCRIPTION• ON/OFF: to turn the meter on and off.• ZERO/CFM: to zero the meter prior to measurement, to confirm edited values or t
9ERRORS AND WARNINGSThe instrument shows clear messages when erroneous condition appears. Messages are also displayedwhen the obtained values are outs