Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.
After a few seconds, the display will
show “-0.0-”. The meter is now zeroed
and ready for measurement.
Remove the cuvette.
Add 10 drops of HI 93706A-0 Phosphorus
Reagent A to the cuvette.
Add the content of one packet of
HI 93706B-0 Phosphorus Reagent B to
the cuvette. Replace the cap and shake
gently until dissolution is complete.
Replace the cuvette into the cuvette holder
and ensure that the notch on the cap is
positioned securely into the groove.
Press and hold READ/ /TIMER for three
seconds. The display will show the
countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.