Hanna-instruments HI 96704 Manuel d'utilisateur Page 14

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 24
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 13
14
Remove the blank.
Insert the cuvette with the reacted sample
into the cuvette holder and ensure that
the notch on the cap is positioned securely
into the groove.
Press READ/ /TIMER and the lamp,
cuvette and detector icons will appear on
the display, depending on the
measurement phase.
At the end of measurement, the instrument
directly displays concentration in µg/L of
hydrazine on the LCD.
INTERFERENCES
Interference may be caused by:
Highly colored samples
Highly turbid samples
Aromatic amines
Presence of oxidizing agents will reduce the concentration of hydrazine.
After a few seconds, the display will
show “-0.0-”. The meter is now zeroed