14
Replace the cuvet into the holder and ensure
that the notch on the cap is positioned securely
into the groove.
Hold READ/TIMER for three seconds. The
display will show the countdown prior to
measurement.
Alternatively, wait for 3 minutes and 30 seconds
and just press READ/TIMER.
In both cases “SIP” will blink during
measurement.
The instrument directly displays concentration
in mg/L of ammonia nitrogen (NH
3
-N) on the
Liquid Crystal Display.
To convert the reading to mg/L of ammonia
(NH
3
), multiply by a factor of 1.216.
or
INTERFERENCES
Organic compounds like: chloramines, various aliphatic and aromatic amines, glycine,
(positive error).
To eliminate these interferences distillation is required.
Organic compounds like: aldehydes, alcohols (e.g. ethanol) or acetone (negative error).
To eliminate these interferences distillation is required.
Sulfide (S
2-
): may cause turbidity.
Hardness above 1 g/L as Calcium Carbonate (CaCO
3
).