The LCD will show "Err5" to indicate that the electrode is now
ready for slope calibration.
Immerse the pH electrode and the probe into
pH4.01 or 10.01 solution.
The LCD will show the corresponding buffer value
at the working temperature.
Wait approximately 30 seconds, and then press
the CON key to confirm the calibration.
The printer will confirm the calibration process
is complete
The calibration is now complete.
4 cm
(1½
")
HI 7004
ºC
pH
ºC
pH
°C
°C
Calibration com-
pleted
CON