Hanna-instruments HI 84431 Manuel d'utilisateur Page 29

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 44
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 28
29
Verify: The instrument has been calibrated (pH and pump) before performing a lot of
titrations. An electrode calibration in at least one point (8.30 pH buffer) is recommended.
Refer to Setup Configuration Menu (see the “Setup” section)
to set up instrument for your measurement.
Sample preparation: For all measurement samples fill
a beaker up to the 50 mL mark with sample. Place the
stir bar into the beaker and then place the beaker in the
appropriate place on the instrument top.
Note: Water samples must be collected and stored in capped
botlles.
Position the electrode holder on the top of the beaker and
secure it by turning clockwise.
Immerse the pH and the temperature electrodes
approximatively 2 cm (0.8”) into the sample to be tested
paying attention not to touch the stir bar. Use O-Rings
provided to secure the pH electrode in holder.
• Insert the dosing tip in the appropriate holder place and
insert it in the sample.
From the titrator main screen press Titration. To enter
titrator main screen from pH meter mode press Titrator
and then Continue.
Press Start to begin the titration process.
TITRATION PROCEDURE
Vue de la page 28
1 2 ... 24 25 26 27 28 29 30 31 32 33 34 ... 43 44

Commentaires sur ces manuels

Pas de commentaire