13
Remove the <0,10 NTU standard cuvet and
place the 15.0 NTU standard cuvet into the
holder.
Close the lid and press READ
. “----” and the
“Lamp and Read Status” indicator will blink
again while making the reading.
At the end of the reading, the third calibration
point (100 NTU) and “CAL P.3” is displayed. If
desired, the value can be changed using the
UP or DOWN keys.
Note: At this moment it is possible to exit
calibration by pressing CAL/RCL. The instrument
will memorize the two-point (<0.10 and 15.0
NTU) calibration data and will return to
measurement mode.
Remove the 15.0 NTU standard cuvet from the
meter and place the 100 NTU standard cuvet
into the holder.
Close the lid and press READ
. “----” and the
“Lamp and Read Status” indicator will blink
again while making the reading.
At the end of the reading, the fourth calibration
point (500 NTU) and “CAL P.4” is displayed. If
desired, the value can be changed using the
UP or DOWN keys.
Note: At this moment it is possible to exit
calibration by pressing CAL/RCL. The instrument
will memorize the three-point (<0.10, 15.0
and 100 NTU) calibration data and will return
to measurement mode.
28
Third calibration point (if available).
Fourth calibration point (if available).
Delete calibration panel.
To delete calibration:
Press SETUP/DEL when the instruments displays
the “Delete calibration” panel.
The user calibration will be deleted and the factory
calibration will be restored. The instrument will
enter automatically in idle mode.