Instruction Manual
HI 96832
Propylene Glycol Refractometer
www.hannainst.com
Vue de la page 0
1 2 3 4 5 6 ... 11 12

Résumé du contenu

Page 1 - HI 96832

1Instruction ManualHI 96832Propylene Glycol Refractometerwww.hannainst.com

Page 2 - TABLE OF CONTENTS

10To make a Propylene Glycol Solution, follow the procedure below:• To make an X % propylene glycol solution volumetrically add X mL of high purity pr

Page 3 - SPECIFICATIONS

11ERROR MESSAGESERROR MESSAGESERROR MESSAGESERROR MESSAGESERROR MESSAGESedoCrorrEDCLnoitpircseD“rrE“otrewopelcyC.eruliaflareneGtcatnoc,stsisreprorrefI

Page 4 - PRINCIPLE OF OPERATION

12Printed in EUROPE(ROMANIA)Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USATechnical Support for CustomersTe

Page 5 - FUNCTIONAL DESCRIPTION

2All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of thecopyright owner, Hanna Instruments Inc., Wo

Page 6 - DISPLAY ELEMENTS

3GENERAL DESCRIPTIONSPECIFICATIONSTemperature Compensation: Automatic between 0 and 40 °C (32 - 104 °F)Measurement Time: Approximately 1.5 secondsMini

Page 7

4Propylene Glycol determinations are made by measuring the refractive index of a solution. RefractiveIndex is an optical characteristic of a substance

Page 8 - MEASUREMENT PROCEDURE

5MEASUREMENT UNITSFreezing point is displayed as a temperature from 0.0 to -51.0 °C corresponding to 0-60 % byvolume.When the display shows “LO”, the

Page 9 - 8 seconds

6DISPLAY ELEMENTSDISPLAY ELEMENTSDISPLAY ELEMENTSDISPLAY ELEMENTSDISPLAY ELEMENTS1. Battery (blinks when low battery condition detected)2. Measurement

Page 10 - 3.6-(C°3.12

7• Handle instrument carefully. Do not drop.• Do not immerse instrument under water.• Do not spray water to any part of instrument except the “sample

Page 11 - ERROR MESSAGES

8MEASUREMENT PROCEDUREMEASUREMENT PROCEDUREMEASUREMENT PROCEDUREMEASUREMENT PROCEDUREMEASUREMENT PROCEDUREVerify the instrument has been calibrated be

Page 12 - MAN96832 01/10

94. Remove sample from the sample well by absorbingwith a soft tissue.5. Using a plastic pipette, rinse prism and samplewell with distilled or deioniz