1. GLP/ key: press to enter
GLP mode
. In
calibration mode
press to
edit the date and time.
2. CAL CHECK key: press to perform the validation of the meter, or
press and hold for three seconds to enter
calibration mode
.
3. ZERO/CFM key: press to zero the meter prior to measurement, to
confirm edited values or to confirm factory calibration restore.
4. READ/ /TIMER key: In
measurement mode
, press to make a
measurement, or press and hold for three seconds to start a
pre-programmed countdown prior to measurement. In
GLP mode
press to view the next screen.
5. ON/OFF key: to turn the meter on and off.
6. Liquid Crystal Display (LCD)
7. Cuvette alignment indicator
8. Cuvette holder
DISPLAY ELEMENTS DESCRIPTION:
INSTRUCTION MANUAL
HI 96730HI 96730
HI 96730HI 96730
HI 96730
www.hannainst.com
Preliminary examination:
Please examine this product carefully. Make sure that the instrument is
not damaged. If any damage occured during shipment, please notify
your Dealer.
Each HI 96730 Ion Selective Meter is supplied complete with:
• Two Sample Cuvettes and Caps
• 9V Battery
• Instruction Manual
Note: Save all packing material until you are sure that the instrument
works correctly. Any defective item must be returned in its original
packing.
For more details about spare parts and accessories
see “Accessories”.
Dear Customer,
Thank you for choosing a Hanna product. This manual will provide you
with the necessary information for the correct use of the instrument. Please
read it carefully before using the meter. If you need additional technical
information, do not hesitate to e-mail us at [email protected].
Functional description:
Light High: There is too much light to perform
a measurement. Please check the preparation of
the zero cuvette.
Light Low: There is not enough light to perform
a measurement. Please check the preparation of
the zero cuvette.
No Light: The instrument cannot adjust the
light level. Please check that the sample does
not contain any debris.
Inverted cuvettes: The sample and the zero
cuvette are inverted.
Zero: A zero reading was not taken. Follow the
instructions of the measurement procedure for
zeroing the meter.
Under range: A blinking “0.00” indicates that
the sample absorbs less light than the zero
reference. Check the procedure and make sure
you use the same cuvette for reference (zero)
and measurement.
Over Range: A flashing value of the maximum
concentration indicates an over range condition.
The concentration of the sample is beyond the
programmed range: dilute the sample and
re-run the test.
Standard Low: The standard reading is less
than expected.
Standard High: The standard reading is higher
than expected.
Errors and warnings:
Measurement procedure:
1• Turn the meter on by pressing ON/
OFF.
2• When the beeper sounds briefly and
the LCD displays dashes, the meter is
ready. The blinking “ZERO” indicates
that the instrument needs to be zeroed
first.
3• Fill the cuvette with 10 mL of
unreacted sample, up to the mark,
and replace the cap.
4• Place the cuvette into the cuvette
holder and ensure that the notch on
the cap is positioned securely into the
groove.
5• Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.
6• After a few seconds the display will
show “-0.0-”. The meter is now zeroed
and ready for measurement. Remove
the cuvette.
7• Fill one graduated mixing cylinder up
to the 25 mL mark with sample.
8• Add the content of one packet of
HI 93730A-0 Molybdenum reagent,
close the cylinder and invert it several
times until the powder is completly
dissolved. This is the blank.
9• Add the content of one packet of
HI 93730B-0 Reagent B to the
cylinder, close it and invert several
times until the powder is completely
dissolved.
10•Add the content of one packet of
HI 93730C-0 Reagent C to the
cylinder, close and shake it vigorously.
11•Fill a cuvette with 10 mL of the reacted
sample up to the 10 mL mark and
replace the cap.
12•Place the cuvette into the cuvette
holder and ensure that the notch on
the cap is positioned securely into the
groove.
Measurement
1. The measuring scheme (lamp, cuvette, detector), appears during
different phases of zero or reading measurement
2. Error messages and warnings
3. The battery icon indicates the charge state of the battery
4. The hourglass appears when an internal check is in progress
5. Status messages
6. The chronometer appears when the reaction timer is running
7. The month, day and date icons appear when a date is displayed
8. Four digit main display
9. Measuring units
10. Four digit secondary display
ON ZERO READING:
ON SAMPLE READING:
DURING CALIBRATION PROCEDURE:
OTHER ERRORS AND WARNINGS:
Cooling lamp: The instrument waits for the
lamp to cool down.
Battery low: The battery must be replaced soon.
Dead battery: This indicates that the battery
is dead and must be replaced. Once this
indication is displayed, normal operation of the
instrument will be interrupted. Change the
battery and restart the meter.
Cap error: Appears when external light enters
in the analysis cell. Assure that the cuvette cap
is present.
Molybdenum
ISM
3
4
5-6
7-10
11
12
Technical specifications:
Range 0.0 to 40.0 mg/L
Resolution 0.1 mg/L
Accuracy ±0.3 mg/L ±5% of reading @ 25°C
Typical EMC Dev. ±0.1 mg/L
Light Source Tungsten Lamp
Light Detector Silicon Photocell with narrow band interference
filter @ 420 nm
Method Adaptation of the mercaptoacetic acid method.
The reaction between molybdenum and the
reagents causes a yellow tint in the sample.
Environment 0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery Type 1 x 9 volt
Auto-Shut off After 10' of non-use in measurement mode;
after 1 hour of non-use in calibration mode;
with last reading reminder.
Dimensions 192 x 104 x 69 mm (7.6 x 4.1 x 2.7 ”)
Weight 360 g (12.7 oz.).
2
Vue de la page 0
1 2

Résumé du contenu

Page 1 - Molybdenum

1. GLP/ key: press to enter GLP mode. In calibration mode press toedit the date and time.2. CAL CHECK key: press to perform the validation of the met

Page 2 - CALIBRATION

11•Then the date of last calibration (e.g.:“01.08.2009”) appears on the display, or“01.01.2009” if the factory calibrationwas selected before. In both