Page 1 - HI 8710 • HI 8711

1www.hannainst.comPanel MountedpH and ORPIndicators andControllersHI 8510 • HI 8512HI 8710 • HI 8711HI 8720Instruction Manual

Page 2

10DOSAGE MODE SWITCH Blinks when the dosing switch is in OFF or ONposition.SWITCHOFF/AUTO/ON To set the dosing mode:• OFF - dosing is disabled• AUTO -

Page 3

111. BNC socket for pH electrode2. Input from amplified electrode3. Connections for Pt100 temperature sensor4. Connections for dosing pump5. Reduc/Oxi

Page 4 - MECHANICAL DIMENSIONS

12ΔALARM To display and set the alarm tolerancepH 7 TEST To verify Offset compensationpH 4 TEST To verify amplifier circuitNote: Each time a key is pr

Page 5

13TrimmersSLOPE For Slope calibrationOFFSET For Offset calibrationΔALARM To set the tolerance of the alarmALK. SETCOARSE To coarsely adjust alkaline s

Page 6

141. BNC socket for pH electrode2. Input from amplified electrode3. Connections for Pt100 temperature sensor4. Connections for dosing pump for acid5.

Page 7

15KeypadSET To set the working point of ORP dosageMEASURE To enter measurement mode and to enablediagnostic testsΔALARM To display and set the alarm t

Page 8

16TrimmersCAL For ORP calibrationΔALARM To display and set the alarm toleranceSET/COARSE To coarsely adjust the setpointSET/FINE To finely adjust the

Page 9

171. BNC socket for ORP electrode2. Input from amplified electrode3. Oxid/Reduc dosage consent terminals4. Connections for dosing pump5. Alarm contact

Page 10

18TUPNIedortcelErettimsnarTEGNAR Vm0001±NOITULOSER Vm1YCARUCCA)F°86/C°02@(Vm5±%5.0±STUPNI01ecnadepmihgiH21;mhOerastupninipgnihctaMdnaecnerefeRelbaliav

Page 11

19TUPNIedortcelErettimsnarTEGNAR Hp00.41ot00.0NOITULOSER Hp10.0YCARUCCA)C°86/C°02@()C°001ot0(Hp20.0±)C°0ot02-(Hp50.0±%5.0±STUPNI01ecnadepmihgiH21;mhOe

Page 12

2PRELIMINARY EXAMINATION ... 3GENERAL DESCRIPTION ...

Page 13 - Switches

20HI 8720TUPNIedortcelErettimsnarTEGNAR Vm0001±NOITULOSER Vm1YCARUCCA)C°86/C°02@(Vm5±%5.0±STUPNI01ecnadepmihgiH21;mhOerastupniniPgnihctaMdnaecnerefeRe

Page 14

21••••• Connect a 3-wire cable to the power supplyterminal according to the voltage level asindicated, and pay particular attention tothe correct line

Page 15

22••••• The HI 8710 models are single dosagecontrollers with acid/alkaline selection.If acid dosage is needed (e.g. in chromiumVI reduction), leave o

Page 16

23••••• CONSENT contacts (HI 8710 and HI 8720,see picture, terminals 3 and 4): these contacts(max. 2A, 240V) are used for reduction andoxidation react

Page 17

24All instrument settings are made via front panel keys and trimmers.When a key is pressed, the corresponding LED lights up to show theoperating funct

Page 18 - SPECIFICATIONS

25SET POINTS (HI 8711)To set the working point for alkaline dosage, press the ALK. SET key andthe display will show the set value for alkaline dosage.

Page 19

26Examples:For HI 8710, if the set value is pH 3 and the ΔALARM is 1.5 pH, theinstrument generates an alarm every time the pH reading is higherthan 4.

Page 20

27When acid dosage is active, the ACID LED lightsup, while during alkaline dosage, the ALK. LEDturns on (HI 8710 only).When oxidant dosage is active,

Page 21 - INITIAL PREPARATION

28Make sure that the instrument is in measurementmode (MEASURE LED is on) before proceedingwith calibration.Measure the temperature of the calibration

Page 22

29Rinse pH electrode and the reference thermometer thoroughly with pH4.01 rinsing solution, then immerse them in pH 4.01 (HI 7004) or pH10.01 (HI 7010

Page 23

3Remove the instrument from the packing material and examine itcarefully to make sure that no damage has occurred during shipping.If there is any noti

Page 24 - OPERATIONAL GUIDE

30Temperature has an effect on the pH. The calibration buffer solutionsare affected by temperature changes to a lesser degree than normalsolutions.Ple

Page 25

31HI 8510, HI 8710 and HI 8711 are provided with autodiagnosticfunctions that allow to check and troubleshoot any malfunctioning.The functions are mad

Page 26

32C)Amplifier Circuit TestPress the pH 4 TEST key and the display should show a valuewithin the 3.30 to 4.30 pH range, to verify the amplifier circuit

Page 27

33HI 8512 and HI 8720 are ORP controllers provided with autodiagnosticfunctions that allow to check and troubleshoot any malfunctioning.The functions

Page 28

34All LEDs above the keys indicate the state of each function, whether itis active or the display is indicating the mode.For HI 8711 onlyEach LED can

Page 29

35Redox measurements allow the quantification of the solution oxidizing/reducing power, and are commonly expressed in mV.Oxidation may be defined as t

Page 30

36Oxidizing pre-treatment: immerse the electrode for some minutes inHI 7092 solution.If no pre-treatment is performed, the electrode will have long re

Page 31 - Sensor Test

37PREPARATION PROCEDURERemove the protective cap.DO NOT BE ALARMED IF ANY SALT DEPOSITS ARE PRESENT. This isnormal with electrodes and they will disap

Page 32 - Amplifier Circuit Test

38PERIODIC MAINTENANCEInspect electrode and cable. The cable used for the connection to themeter must be intact and there must be no points of broken

Page 33 - ORP DIAGNOSTIC TESTS

39CLEANING PROCEDURE• General Soak in HI 7061 general cleaning solution forapproximately 1 hour.Removal of films, dirt or deposits on the membran

Page 34 - LED INDICATION

4The meters are provided with a black anodized aluminum body,front and back panels in shockproof ABS plastic and a transparentsplash-proof front cover

Page 35 - TAKING REDOX MEASUREMENTS

40SHORT DISTANCE, INDOOR INSTALLATIONDue to the low current involved, a very high grade of insulation isrequired.A dry environment is needed in order

Page 36

41AmpHel® is a registered Trademark of "Hanna Instruments"LONG DISTANCE INSTALLATIONS, ISOLATED OUTPUT FOR PC INTERFACEIf the needed install

Page 37 - ELECTRODE MAINTENENCE

42pH CALIBRATION SOLUTIONSHI 7004M pH 4.01 buffer solution, 230 mLHI 7004L pH 4.01 buffer solution, 500 mLHI 7006M pH 6.86 buffer solution, 230 mLHI 7

Page 38

43pH ELECTRODESHI 1090T Screwcap PG13.5 connector, double junction, glassbodyHI 1110S Screw connector, single junction, glass bodyHI 1130B/3 BNC conne

Page 39

44HI 1210T Screwcap PG13.5 connector, double junction, plas-tic bodyHI 1910B BNC connector, 1 m (3.3') cable, double junction,plastic body, built

Page 40 - SUGGESTED INSTALLATIONS

45ORP ELECTRODESHI 2930B/5 BNC connector, 5 m (16.5') cable, Pt, plastic body,built-in amplifierHI 3110S Screw connector, Pt, glass bodyHI 3130B

Page 41

46OTHER ACCESSORIESHI 98501ChecktempC thermometer with penetration probeand 0.1°C resolution (-50.0 to 150.0°C)HI 8614 pH transmitter (0.00 to 14.00

Page 42 - ACCESSORIES

47HANNA instruments reserves the right to modify the design, construction andappearance of its products without advance notice.HI 778P Coaxial cable a

Page 43

48Printed in Romania, EU MAN8711 09/09Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USALocal Sales and Custom

Page 44

5FRONT PANELKeypadSENSOR TEST To display the mV reading of the electrode and,therefore, verify its working conditionpH 7 TEST To verify the internal c

Page 45

6Unplug the instrument from the power supply before replacing the fuse.Only one of BNC, AMP INPUT or INPUT TRANSMITTER connectors can beused at a mome

Page 46 - ChecktempC

7Keypad0 mV TEST To verify the instrument calibration at 0 mV1000 mV TEST To verify the slope at 1000 mVNote: Each time a key is pressed the correspon

Page 47

81. BNC socket for ORP electrode2. Input from amplified electrode3. Power supply terminals4. Fuse holder5. Recorder output terminals6. Connection to t

Page 48

9KeypadSET To set the pH dosage limitMEASURE To enter measurement mode and to enablediagnostic testsSENSOR TEST To display electrode mV reading and ve