1Instruction ManualHI 83200Multiparameter BenchPhotometerfor Laboratorieswww.hannainst.comThis Instrument is inCompliance with the CE Directives
18 19• Remove the cap and fill the Demineralizer Bottle with tapwater.• Replace the cap and shake gently for at least 2 minutes.• Open the upper part
20 21AluminumALUMINUMSPECIFICATIONSRange 0.00 to 1.00 mg/LResolution 0.01 mg/LAccuracy ±0.02 mg/L ±4% of readingTypical EMC ±0.01 mg/LDeviationLight S
22 23• Add the content of one packet of HI 93712C-0 Bleachingpowder to one of the two cuvettes. Replace the cap andshake vigorously until completely d
24 25Ammonia MRSPECIFICATIONSRange 0.00 to 10.00 mg/LResolution 0.01 mg/LAccuracy ±0.05 mg/L ±5% of readingTypical EMC ±0.01 mg/LDeviationLight Source
26 27• Remove the cuvette.• Add 4 drops of HI 93715A-0 First Reagent (6 drops for seawateranalysis). Replace the cap and mix the solution.• Add 4 drop
28 29Ammonia LRBromineBROMINESPECIFICATIONSRange 0.00 to 8.00 mg/LResolution 0.01 mg/LAccuracy ±0.08 mg/L ±3% of readingTypical EMC ±0.01 mg/LDeviatio
30 31• Remove the cuvette and add the content of one packet of HI 93716-0DPD reagent. Replace the cap and shake gently for about 20 seconds todissolve
32 33FREE CHLORINEFREE CHLORINEFREE CHLORINEFREE CHLORINEFREE CHLORINESPECIFICATIONSRange 0.00 to 2.50 mg/LResolution 0.01 mg/LAccuracy ±0.03 mg/L ±3%
34 35• Reinsert the cuvette into the instrument.• Press READ to start the reading. The instrument displays the resultsin mg/L of free chlorine.INTERFE
36 37Total Chlorine• Press ZERO key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.• Remove the cuvette.Powder reagen
2 3PRELIMINARY EXAMINATIONPlease examine this product carefully. Make sure that the instrument is not damaged. If any damageoccurred during shipment,
38 39Chlorine DioxideCHLORINE DIOXIDESPECIFICATIONSRange 0.00 to 2.00 mg/LResolution 0.01 mg/LAccuracy ±0.10 mg/L ±5% of readingTypical EMC ±0.01 mg/L
40 41Chlorine Dioxide• Press ZERO key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.• Fill another cuvette with 10 m
42 43• Reinsert the cuvette into the instrument.• Press TIMER and the display will show the countdown prior to themeasurement or, alternatively, wait
44 45• Reinsert the cuvette into the instrument.• Press TIMER and the display will show the countdown prior to themeasurement or, alternatively, wait
46 47Color of Water# 3• Fill the second cuvette up to the mark with unfiltered sample andreplace the cap. This is the apparent color.• Filter 10 mL of
48 49Copper HR• Remove the cuvette.• Add the content of one packet of HI 93702-0 Bicinchoninate. Replacethe cap and shake gently for about 15 seconds.
50 51Copper LR• Remove the cuvette.• Add the content of one packet of HI 93747-0 Bicinchoninate.Replace the cap and shake gently for about 15 seconds.
52 53CyanideNote: Pay attention to the way the spoon is filled:- do not press the powder;- do not overfill it.• Place the HDPE plastic stopper and cap
54 55Cyanuric AcidCYANURIC ACIDSPECIFICATIONSRange 0 to 80 mg/LResolution 1 mg/LAccuracy ±1 mg/L ±15% of readingTypical EMC ±1 mg/LDeviationLight Sour
56 57FluorideFLUORIDESPECIFICATIONSRange 0.00 to 2.00 mg/LResolution 0.01 mg/LAccuracy ±5% of readingTypical EMC ±0.01 mg/LDeviationLight Source Tungs
4 5Where:-log I/Io=Absorbance (A)Io=intensity of incident light beamI =intensity of light beam after absorptionελ=molar extinction coefficient at wave
58 59Hardness CaCALCIUM HARDNESSSPECIFICATIONSRange 0.00 to 2.70 mg/LResolution 0.01 mg/LAccuracy ±0.11 mg/L ±5% of readingTypical EMC ±0.01 mg/LDevia
60 61Hardness CaNote: This test will detect any calcium contamination in the beaker, measuring syringes or sample cells. Totest cleanliness, repeat th
62 63Hardness Mg• Fill both cuvettes up to the 10 mL mark.• Add 1 drop of HI 93719C-0 EDTA solution to one cuvette (# 1),replace the cap and invert th
64 65HydrazineHYDRAZINESPECIFICATIONSRange 0 to 400 µg/LResolution 1 µg/LAccuracy ±4% of full scaleTypical EMC ±2 µg/LDeviationLight Source Tungsten l
66 67IODINESPECIFICATIONSRange 0.0 to 12.5 mg/LResolution 0.1 mg/LAccuracy ±0.1 mg/L ±5% of readingTypical EMC ±0.1 mg/LDeviationLight Source Tungsten
68 69Iron HRIRON HIGH RANGESPECIFICATIONSRange 0.00 to 5.00 mg/LResolution 0.01 mg/LAccuracy ±0.04 mg/L ±2% of readingTypical EMC ±0.01 mg/LDeviationL
70 71Iron LRIRON LOW RANGESPECIFICATIONSRange 0 to 400 µg/LResolution 1 µg/LAccuracy ±10 µg/L ±8% of readingTypical EMC ±1 µg/LDeviationLight Source T
72 73Iron LRINTERFERENCESInterference may be caused by:Cadmium above 4.0 mg/LChromium3+ above 0.25 mg/LChromium6+ above 1.2 mg/LCobalt above 0.05 mg/L
74 75Magnesium• Remove the cuvette.• Using the other 1 mL syringe, add to the cuvette exactly 0.5 mL of sample.Note: Do not mix up the two syringes!•
76 77Manganese HR• Add the content of one packet of HI 93709B-0 Sodium Periodatereagent. Replace the cap and shake gently until completely dissolved.•
6 7INSTRUMENT DESCRIPTION1) Cuvette Lid2) Liquid Crystal Display (LCD).3) Splash proof keypad.4) ON/OFF power switch5) Power input connector6) USB con
78 79Manganese LR• Add 0.2 mL of the HI 93748B-0 Alkaline-cyanide reagentsolution to each cuvette, replace the caps and shake gently.• Add 1 mL of the
80 81 MolybdenumMOLYBDENUMSPECIFICATIONSRange 0.0 to 40.0 mg/LResolution 0.1 mg/LAccuracy ±0.3 mg/L ±5% of readingTypical EMC ±0.1 mg/LDeviationLight
82 83• Fill one graduated mixing cylinder up to the 25 mLmark with the sample.• Add the content of one packet of HI 93730A-0molybdenum reagent, close
84 85Nickel HR• Reinsert the cuvette into the instrument.• Press TIMER and the display will show the countdown priorto the measurement or, alternative
86 87Nickel LR• Press TIMER and the display will show a countdown prior to adding reagent C or, alternatively, waitfor 15 minutes. Add one packet of H
88 89NitrateSPECIFICATIONSRange 0.0 to 30.0 mg/LResolution 0.1 mg/LAccuracy ±0.5 mg/L ±10% of readingTypical EMC ±0.1 mg/LDeviationLight Source Tungst
90 91Nitrite HRNITRITE HIGH RANGESPECIFICATIONSRange 0 to 150 mg/LResolution 1 mg/LAccuracy ±4 mg/L ±4% of readingTypical EMC ±1 mg/LDeviationLight So
92 93SPECIFICATIONSRange 0.00 to 0.35 mg/LResolution 0.01 mg/LAccuracy ±0.02 mg/L ±4% of readingTypical EMC ±0.01 mg/LDeviationLight Source Tungsten l
94 95Dissolved OxygenSPECIFICATIONSRange 0.0 to 10.0 mg/LResolution 0.1 mg/LAccuracy ±0.4 mg/L ±3% of readingTypical EMC ± 0.1 mg/LDeviationLight Sour
96 97OZONESPECIFICATIONSRange 0.00 to 2.00 mg/LResolution 0.01 mg/LPrecision ±0.03 mg/L @ 1.00 mg/LTypical EMC ±0.01 mg/LDeviationLight Source Tungste
8 9USING LIQUID AND POWDER REAGENTS• Proper use of the dropper:(a) for reproducible results, tap the dropper on the table for several times and wipe
98 99• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.• Remove the cuvette.• Add the content of one
100 101pH• Reinsert the cuvette into the instrument.• Press the READ key to start the reading. The instrument displaysthe pH value.pHpHSPECIFICATIONSR
102 103Phosphate HR• Add the content of one packet of HI 93717B-0 PhosphateHR Reagent B to the cuvette. Replace the cap and shakegently until complete
104 105Phosphate LR• Reinsert the cuvette into the instrument.• Press TIMER and the display will show the countdown prior to themeasurement or, altern
106 107Phosphorus• Add the content of one packet of HI 93706B-0 PhosphorusReagent B (Amino Acid) to the cuvette. Replace the capand shake gently until
108 109• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.• Remove the cuvette and add the content of
110 111Potassium MRPOTASSIUM MEDIUM RANGESPECIFICATIONSRange 10 to 100 mg/LResolution 2.5 mg/LAccuracy ±15 mg/L ±7% of readingTypical EMC ±2.5 mg/LDev
112 113Potassium LRPOTASSIUM LOW RANGESPECIFICATIONSRange 0.0 to 20.0 mg/LResolution 0.5 mg/LAccuracy ±3.0 mg/L ±7% of readingTypical EMC ±0.5 mg/LDev
114 115• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.• Remove the cuvette and add the content of
116 117Silica• Place the cuvette into the holder and close the lid.• Press ZERO key. The display will show “-0.0-” when themeter is zeroed and ready f
10 11Pay attention to push the cuvette completely down in the holder.• In order to avoid reagent leaking and to obtain more accurate measure-ments, cl
118 119Silver• Add exactly 1.0 mL of HI 93737B-0 Buffer reagent to the second beaker(the sample) and swirl gently to mix.• Press TIMER and the display
120 121SPECIFICATIONSRange 0 to 150 mg/LResolution 5 mg/LAccuracy ±5 mg/L ±3% of readingLight Source Tungsten lamp with narrow band interference filte
122 123ZincZINCSPECIFICATIONSRange 0.00 to 3.00 mg/LResolution 0.01 mg/LAccuracy ±0.03 mg/L ±3% of readingTypical EMC ±0.01 mg/LDeviationLight Source
124 125The instrument shows clear warning messages when erroneous conditions appear and when measured values areoutside the expected range. These mess
126 127Description Range MethodAluminum 0.00 to 1.00 mg/L AluminonAlkalinity 0 to 500 mg/L ColorimetricAmmonia MR 0.00 to 10.00 mg /L NesslerAmmonia L
128 129HI 93721-01 100 iron HR testsHI 93721-03 300 iron HR testsHI 93722-01 100 cyanuric acid testsHI 93722-03 300 cyanuric acid testsHI 93723-01 100
130 131WARRANTYAll Hanna Instruments meters are warranted for two years against defects in workmanship and materialswhen used for its intended purpos
132Australia:Tel. (03) 9769.0666 • Fax (03) 9769.0699e-mail: [email protected]:Tel. (10) 88570068 • Fax (10) 88570060e-mail: hannachina@v
12 13POWER CONNECTION AND BATTERY MANAGEMENTThe meter can be powered from an AC/DC adapter (included) or from the built-in rechargeable battery.Note:
14 15Date / TimeThis option is used to set the instrument’s date and time.Press Modify functional key to change the date/time.Press the WX functional
16 17HI 83200 offers an interactive contextual help mode that assists the user at any time.To access help screens press HELP.The instrument will displ