Hanna-instruments HI 9321 Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Instruments de mesure Hanna-instruments HI 9321. Hanna Instruments HI 9321 User Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 36
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 0
Instruction Manual
HI 9321
HI 931400
HI 931401
Microprocessor-based
Bench pH/mV/°C Meters
www.hannainst.com
These Instruments are in
Compliance with the CE Directives
O
FF
°
C
O
N
M
R
°
C
M
E
M
R
A
N
G
E
C
F
M
C
O
M
M
C
A
L
H
I 9
3
1
4
0
1
M
IC
R
O
P
R
O
C
E
S
S
O
R
p
H
M
E
T
E
R
CFM
Confirm when ready
/
Select buffer 2 value
CFM
Confirm when ready
/
Select buffer 1 value
CAL
Start calibration
p
H
c
a
lib
r
a
tio
n
p
ro
c
e
d
u
r
e
RANGE
Select pH or ORP measurement
COM
M
For RS232 configuration
M
R
Recall previous stored data
M
EM
For storing display result
F
u
n
c
tio
n
°
C
p
H
ins
rum
n
s
Vue de la page 0
1 2 3 4 5 6 ... 35 36

Résumé du contenu

Page 1 - Bench pH/mV/°C Meters

Instruction ManualHI 9321HI 931400HI 931401Microprocessor-basedBench pH/mV/°C Meterswww.hannainst.comThese Instruments are inCompliance with the CE Di

Page 2 - TABLE OF CONTENTS

10In order to take accurate pH measurements, make sure that theinstrument has been calibrated before use (see page 12).For more accurate results, it i

Page 3 - GENERAL DESCRIPTION

11ORP MEASUREMENTSPress RANGE.Submerge the ORP or ISE electrode tip (4cm/1½") into the sample.Allow a few minutes for the reading to stabi-lize.I

Page 4 - 23 4 5 6 7 8

12RESET BUTTONThe RESET button (see pages 4-5, #V) should be used when theinstrument displays erroneous messages due to strong electrical inter-ferenc

Page 5 - HI 931400 & HI 931401

13PROCEDURECalibration has a choice of 5 memorized buffers: 4.01, 6.86, 7.01,9.18 and 10.01 pH.It is always recommended to perform a two-point calibra

Page 6 - & HI 931400

14close to the selected buffer, the meterstores the reading. The buffer value isthen displayed on the primary LCD andthe secondary LCD will display &q

Page 7 - SPECIFICATIONS HI 931401

15Note: The meter will automatically skip the buffer that was used forthe first calibration. pH 6.86 and pH 7.01 as well as 9.18and 10.01 pairs of buf

Page 8 - HI 9231 & HI 931400 LCD

16pH VALUES AT VARIOUSTEMPERATURESTEMP pH VALUES°C °F 4.01 6.86 7.01 9.18 10.01051015202530354045505560657032415059687786951041131221311401491584.014.

Page 9 - OPERATIONAL GUIDE

17TEMPERATURE AND mV CALIBRATIONThe pH meter has been factory calibrated for temperature and mV.The temperature probes and ORP electrodes are intercha

Page 10

18TROUBLESHOOTING GUIDESymptoms Problem SolutionThe meter is slow inresponding or givesfaulty readingsThe meter does notaccept the 2nd buffersolution

Page 11 - HOLD FUNCTION

19ELECTRODE CONDITIONING& MAINTENANCEPREPARATIONRemove the protective cap.DO NOT BE ALARMED IF ANY SALT DEPOSITS ARE PRESENT.This is normal with e

Page 12 - PREPARATION

2Dear Customer,Thank you for choosing a Hanna Product. It is recommended that youread this instruction manual carefully before using the instrument.Th

Page 13 - PROCEDURE

20For refillable electrodes:If the filling solution (electrolyte) is more than 2½ cm (1") below thefill hole, add HI 7082 or HI 8082 3.5M KCl Ele

Page 14

21CLEANING PROCEDUREGeneral Soak in Hanna HI 7061 or HI 8061 General CleaningSolution for approximately ½ hour.Removal of films, dirt or deposits on t

Page 15

22Since the resistance of the pH electrode is in the 200 Mohm range,the current across the membrane is in the pico Ampere range. Largecurrents can dis

Page 16 - TEMPERATURES

23High concentrations of sodium ions interfere with readings in alkalinesolutions. The pH at which the interference starts to be significantdepends up

Page 17 - INTERFACE WITH PC

24ADDITIONAL CALIBRATIONPROCEDURES & INFORMATION(for technical personnel only)TEMPERATURE CALIBRATIONInitial Preparation• Prepare a vessel contain

Page 18 - Symptoms Problem Solution

25• When the temperature reading has stabi-lized, use the ↑↑↑↑↑°C or ↓↓↓↓↓°C keys to set thereading on the secondary LCD to that ofthe reference therm

Page 19 - & MAINTENANCE

26The following baud rate can be selected with↑↑↑↑↑°C and ↓↓↓↓↓°C: 150, 300, 600, 1200 (factorysetting) and 2400.Press CFM to confirm the setting and

Page 20 - PERIODIC MAINTENANCE

27DWC is equivalent to pressing the ↓↓↓↓↓°C keyMEM is equivalent to pressing the MEM keyMRR is equivalent to pressing the MR keyCOM is equivalent to p

Page 21 - CLEANING PROCEDURE

28ELECTRODE STORAGE SOLUTIONHI 70300L Storage Solution, 460 mLELECTRODE STORAGE SOLUTION IN FDA APPROVEDBOTTLESHI 80300L Storage Solution, 460 mLELECT

Page 22 - FOR HANNA pH SENSITIVE GLASS

29pH & ORP ELECTRODESAll electrodes part numbers ending in B are supplied with a BNCconnector and 1 m (3.3') cable.HI 1043BGlass-body, double

Page 23

3PRELIMINARY EXAMINATIONThe Hanna Instruments HI 9321, HI 931400 and HI 931401 aremicroprocessor-based bench meters for pH, mV (ORP and ISE) andtemper

Page 24 - PROCEDURES & INFORMATION

30HI 1330BGlass-body, semimicro, single junction, refillable, combination pHelectrode. Use: laboratory, vials.HI 1331BGlass-body, semimicro, single ju

Page 25

31FC 100BPlastic-body (Kynar®), double junction, refillable, combination pHelectrode. Use: general purpose for food industry.FC 200BPlastic-body (Kyna

Page 26 - Windows Terminal

32FC 911BPlastic-body (Kynar®), double junction, refillable with built-in ampli-fier, combination pH electrode. Use: very high humidity.HI 1413BGlass-

Page 27 - ACCESSORIES

33HI 5412Glass-body, single Calomel reference half-cell electrode, refillablewith 4mm plug with 1m (3.3') cable. Use: general purpose withconsta

Page 28

34ACCESSORIESHI 710005 Transformer 110VAC to 12 VDCHI 710006 Transformer 220VAC to 12 VDCChecktempC Pocket-size thermometer (range -50.0 to 150.0°C)H

Page 29 - "S" VERSION

35CE DECLARATION OF CONFORMITYHanna Instruments reserves the right to modify the design,construction and appearance of its products without advancenot

Page 30

www.hannainst.comMAN9321R2 03/97

Page 31 - 0.25"

4FUNCTIONAL DESCRIPTION HI 9321MR°CMEMON/OFF°CRANGECFMCALCFM Confirm when ready/ Select buffer 2 valueCFM Confirm when ready/ Select buffer 1 valueCAL

Page 32 - 0.5"

5I. BNC SocketII. Reference Electrode SocketIII. Reference Electrode SocketIV. Temperature Probe SocketV. Reset ButtonVI. VDC Power Socket (for HI 710

Page 33 - 4.7"

6SPECIFICATIONS HI 9321& HI 931400Range pH 0.00 to 14.00ISE ±399.9 mVORP ±1999 mV°C 0.0 to 100.0Resolution pH 0.01 ISE 0.1 mVORP 1 mV°C 0.1Accurac

Page 34 - WARRANTY

7SPECIFICATIONS HI 931401Range pH 0.00 to 14.00 / 0.000 to 14.000mV ±999.9 (ISE & ORP)°C 0.0 to 100.0Resolution pH 0.01 / 0.001 mV 0.1°C 0.1Accura

Page 35 - CE DECLARATION OF CONFORMITY

8FUNCTIONAL DESCRIPTION OFHI 931401 LCDPrimary DisplaySecondary Display°C12pHpHmVCALMEMNOT READYCON WRONGLOW BATBUFFUNCTIONAL DESCRIPTION OFHI 9231 &

Page 36 - MAN9321R2

9OPERATIONAL GUIDEELECTRODE AND PROBE CONNECTIONSFor combination pH or ORP electrodes (with internal reference) plugthe electrode's BNC to the so